Internal jugular access using pocket ultrasound in a simulated model: comparison between biplane and monoplane visualization techniques

Abstract

Introduction

Ultrasound is the current standard for central venous access due to its advantages in efficiency and safety. In-plane and out-of-plane visualization techniques are commonly used, but there is no clear evidence showing an advantage of one technique over the other. The objective of this study was to compare the success and time required for biplane visualization vs. in-plane and out-of-plane techniques in simulated models.

Methodology

Ten emergency medicine specialists participated in 60 simulated events, with randomization of the visualization technique for each event. Each event required intravenous cannulation of a simulated model for jugular venous access, with a maximum of three attempts allowed. The number of attempts required for each event, success of puncture and venous cannulation, frequency of redirection and puncture of the posterior wall, time required to obtain an optimal window, visualize the needle inside the vessel, and passage of the guidewire were recorded. The success ratios and times required for each visualization technique (biplane, in-plane, and out-of-plane) were compared.

Results

Cannulation success rate was 100% for all three techniques. Success on the first attempt was 95% for biplane visualization vs. 100% for in-plane and out-of-plane. The median total time for the procedure was higher for biplane visualization (29.9 s) compared to in-plane (25.2 s) and out-of-plane (29 s), but this difference was not statistically significant (p = 0.999). There were no significant differences in cannulation success, needle redirection, or posterior wall puncture frequency between biplane visualization and in-plane and out-of-plane techniques.

Conclusions

This study suggests that biplane visualization with the use of pocket ultrasound for internal jugular cannulation in simulated models did not demonstrate significant differences when compared with in-plane and out-of-plane visualization techniques. Further research with larger sample sizes may be needed to confirm these results.

Introduction

Central venous access is a common procedure used for patients both inside and outside of intensive care units to insert various devices for patient hemodynamic monitoring, pharmacological and hydroelectrolytic therapies, renal replacement therapies, parenteral nutrition, and cardiac stimulation, among others [1]. In the United States, more than 5 million central venous catheters are inserted each year [2, 3]. Central catheter use is higher in the ICU (55.4%) than outside the ICU (24.4%) [4], representing a significant cost to the healthcare system of 17–29 billion dollars per year [5].

This intervention is associated with thrombotic, infectious, and mechanical complications, with reported frequencies of 15–26% [3, 6]. However, the use of ultrasound as a guide for insertion visualization has led to a decrease in complication rates in recent years. Systematic reviews have shown a decrease in complication rates of up to 71% with the use of ultrasound [7], resulting in a higher success rate and a complication rate of 4.6% [8]. Major medical associations now recommend its use [9,10,11,12].

The in-plane and out-of-plane techniques are the most frequent techniques used for ultrasound-guided central venous access. However, no difference has been found between them [13,14,15]. Recently, the biplane technique has been reported as another technique that combines the two techniques mentioned above for the passage of central catheters [16,17,18]. This study aims to compare the biplane technique vs. in-plane and out-of-plane techniques for the insertion of jugular central venous accesses using a pocket ultrasound machine.

Methodology

This is a single-factor experimental study conducted in simulation with a quantitative approach, developed at the Advanced Center for Clinical Simulation and Technological Innovation of the Universidad del Rosario in Bogota, Colombia. Specialists in emergency medicine with experience in ultrasound-guided vascular access were recruited to perform the punctures using visualization techniques in one and two planes. The call was kept open until 10 specialists who met the inclusion criteria were recruited and subsequently agreed to participate under informed consent. Based on previous studies, cannulation success in the first attempt was estimated at 90% for the biplane technique and 50% for the in-plane and out-of-plane techniques [1], requiring 60 procedures (20 for each technique) to achieve a power of 80% and an α error of 0.05, allowing detection of the difference between the ultrasound windows and the outcomes under study. The arcosine approximation was used, employing the Granmo application version 7.12 April 2012.

Prior to each set of three events, a random assignment of the three ultrasound planes was conducted to determine which one would be used. The randomization was performed using the statistical package R Core Team 4.2.2.

The procedure was performed on a simulated model for central vascular accesses, CentraLineMan® System CML 50, which had external anatomical landmarks include clavicle, sternal notch, sternal and clavicular heads of the sternocleidomastoid muscle, manubrium, and lateral border of the first rib. The model was placed at a distance of 1.2 cm from the skin to the vessel at the lower third of the neck. A right jugular vein with a diameter of 1 cm and a right carotid artery located 7 mm away from the vein were used. The procedure utilized an adult central venous access device, Certofix® Duo S 720, with an 18G × 2 ¾" needle, 0.89 mm diameter guide, and a length of 50 cm (Fig. 1).

Fig. 1
figure 1

Ultrasound planes for vascular access. A Biplane, B Out-of-plane, C In-plane

Sociodemographic variables such as age and sex, experience-related variables, and procedure-related variables (time to window, time for internal jugular vein puncture, puncture success, posterior wall puncture, puncture success, redirection, and the number of punctures) were included.

The study was conducted in a controlled environment at a university simulation center. To ensure that participants were able to familiarize themselves with the ultrasound machine without becoming familiar with the characteristics of the actual target vessel, a different simulated model of the central venous access (Blue PhantomTM Central Venous Access) was used during the 15-min equipment familiarization period. This was done prior to the procedure, after each participant received a 6-min video explanation of the in-plane, out-of-plane, and biplane central venous access techniques. A researcher was present during this time to clarify any doubts about the operation of the ultrasound machine.

To perform the ultrasound windows, the Butterfly iQ + (pocket ultrasound scanner, reference 900-20006-01) was connected to an iPad Pro (9.7 inches) using the Butterfly Network V2.12.1 software, configured in the vascular access preset, with a depth of 3 cm and a gain of 90%. The participant could make adjustments considered pertinent to obtain real-time visualization of the best ultrasound window of the vessel to be cannulated.

The data were recorded immediately and stored digitally on university servers provided for this purpose (SharePoint-Office 365 license Universidad del Rosario), which only researchers had access to.

Analysis

Categorical variables, such as gender, ultrasound plane, puncture success, cannulation success, redirection, and posterior wall puncture, were described using absolute and relative frequencies. Continuous variables, such as age, years of experience, specific experience, time to window, and time to vessel, were described using means and standard deviations or medians and interquartile ranges, depending on the distribution of the data. The normality of the data was explored using Shapiro–Wilk test.

To determine whether there were significant differences in proportions between the different ultrasound planes (biplane, in-plane, and out-of-plane), the success of cannulation, the success of puncture and posterior wall puncture, and the need to redirect the needle, Fisher's exact test was performed. To determine whether there were significant differences in medians (according to normality) between the different ultrasound planes and the time to window, time to vessel, and number of punctures, the Kruskal–Wallis test was applied. For cannulation time, the one-factor ANOVA test was used after checking the assumptions. Post hoc tests were used to further explore significant differences. A statistical significance level of 0.05% and a 95% confidence interval were used for all tests. The statistical analysis was performed using Jamovi software version 2.2.5.

Results

Ten expert specialists in emergency medicine with experience in ultrasound-guided vascular access were included in the study. Each expert performed a total of six procedures, two for each ultrasound plane evaluated (biplane, in-plane, and out-of-plane). Eighty percent of participants were men, with a mean age of 36 years (RIQ 6.25) and a mean of specialist experience was 5 years (RIQ 2.5).

Successful cannulation of the internal jugular vein

Successful cannulation of the internal jugular vein was achieved with all three techniques at 100%. The puncture success in the first attempt was 95% (n = 19/20) with biplane imaging and 100% (n = 20/20) with single-plane imaging (p = 0.362). The total time taken for each attempt was less with in-plane imaging (27.61 ± 8.96), followed by out-of-plane (30.63 ± 8.81) and biplane (31.45 ± 13.59; p = 0.327) (Fig. 2). No statistically significant differences were found when comparing the total procedure time, the time required to obtain the best ultrasound window, and the time required for puncture among the experts who performed the procedures.

Fig. 2
figure 2

Time to cannulation by ultrasound imaging. The total time in seconds to internal jugular vein cannulation by ultrasound image was not significantly different between the three techniques (p = 0.458)

Regarding the time required to obtain the correct ultrasound window of the internal jugular vein in the first attempt, it was shorter with the out-of-plane image (7.91 RIQ: 4.72), followed by biplane (7.13 RIQ: 4.13) and in-plane (7.25 RIQ: 5.75; p = 0.944), but no statistically significant differences were found (Table 1). The time taken from the puncture of the model surface to correct visualization of the needle inside the vessel was shorter in in-plane (5.49 RIQ: 5.49), followed by out-of-plane (6.23 RIQ: 6.23) and biplane (9.12 RIQ 9.55; p = 0.187). Puncture success was 100% with in-plane and out-of-plane images and 95% in biplane (p = 0.999).

Table 1 Results of outcomes according to ultrasound imaging

There was only one failure in the first attempt when the puncture was not successful with the biplane technique. In the second attempt, the procedure was successful with a window time of 3.68 s, time to vessel of 15.74 s, and a total time of 29.89 s; there was no redirection or puncture of the posterior wall.

Posterior wall puncture and redirection

With in-plane imaging, there were fewer punctures of the posterior wall during the procedure, followed by biplane and out-of-plane (10% vs. 15% vs. 20% p = 0.368), and redirections occurred more frequently in biplane, followed by in-plane and out-of-plane (25% vs. 15% vs. 5% p = 0.265) (Fig. 3).

Fig. 3
figure 3

Puncture of the posterior wall according to ultrasound plane

Discussion

According to the searches performed in different scientific databases by the researchers, this is the first study investigating the performance of a pocket ultrasound machine for internal jugular vein cannulation in a simulated model, comparing biplane visualization not only with out-of-plane visualization but also with in-plane visualization. The results indicate that contrary to the hypothesis of advantage with the biplane technique, no statistically significant differences were found in terms of success, execution time, frequency of posterior wall puncture, or need for needle redirection.

When comparing our study to the one conducted by Convissar et al., which used a simulated model for peripheral access with a smaller diameter compared to the larger diameter model used in our study, differences in the use of biplane vs. out-of-plane ultrasound techniques can be explained [18]. It is possible that the smaller diameter of the peripheral vessels in Convissar's study made visualization and access more difficult, which could have affected the effectiveness of the out-of-plane technique and improved the effectiveness of the biplane technique. In contrast, in our study, the larger diameter of the central vessels may have allowed for easier visualization and access, which could have reduced the differences in the effectiveness of the ultrasound techniques used.

In the study by Scholten et al., where the same type of ultrasound machine was used to compare the biplane technique vs. in-plane and out-of-plane techniques, the choice between any of the 2D techniques was subject to the preference of the professional at the time of performing radial artery catheterization in patients undergoing elective cardiothoracic surgery. It was found that 90% of the professionals in the 2D technique group preferred the use of the out-of-plane vs. in-plane technique [19]. In terms of overall success and success in the first attempt, as well as in the posterior wall puncture, no significant differences were found. No difference was observed in secondary outcomes such as the number of punctures, total procedure time, and perceived mental effort during the procedure by the operator.

In the study by Jones et al., where the biplane vs. out-of-plane technique was compared for the insertion of an internal jugular catheter in a simulated model, a procedure performed by emergency medicine residents with previous experience in the use of ultrasound guidance for obtaining vascular access, no significant differences were found in the success of cannulation or the frequency of puncture of the posterior wall [20]. Regarding secondary outcomes, a significant difference was observed in the time required for cannulation, being shorter in out-of-plane than in biplane. Concerning the success of cannulation on the first attempt or puncture of the posterior wall, no significant differences were found between the techniques.

Limitations

This study has some limitations that should be acknowledged. First, the study was conducted in a simulation model, and the results may not necessarily translate to real patients. Although the simulation model allowed for standardized conditions, it is important to note that the anatomy and physiology of a simulated model may differ from that of a real patient. Therefore, caution should be exercised when interpreting the results of this study and further research should be conducted in real patient populations.

Second, all procedures were performed by emergency physicians with varying degrees of experience. While this reflects the current clinical practice, it may limit the generalizability of the results to other specialties or levels of experience. Future studies should include physicians from different specialties and levels of experience to better understand the applicability of these techniques in various clinical settings.

Despite these limitations, this study provides valuable insights into the efficacy of different ultrasound techniques for vascular access in a simulated model. Further research is needed to confirm these findings in real patient populations with a wider range of healthcare providers.

Conclusions

In this simulated model experiment for ultrasound-guided jugular vascular access with pocket ultrasound operated by emergency medicine physicians, no significant differences were evident when comparing biplane visualization with in-plane or out-of-plane visualization techniques in terms of cannulation success, time required for access, need for needle redirection, or frequency of puncture of the posterior vessel wall. The study results show that all three ultrasound planes performed well not only in terms of success but also in safety, which is important given their much lower cost compared to conventional equipment. These findings have important implications for clinical practice and highlight the potential benefits of using pocket ultrasound machines for internal jugular vein cannulation.

Availability of data and materials

The data stored digitally on the university servers provided for this and are available from the corresponding author on reasonable request.

References

  1. Smith RN, Nolan JP (2013) Central venous catheters. BMJ 11(347):f6570–f6570

    Article  Google Scholar 

  2. Moore CL (2014) Ultrasound first, second, and last for vascular access. J Ultrasound Med 33(7):1135–1142

    Article  PubMed  Google Scholar 

  3. McGee DC, Gould MK (2003) Preventing complications of central venous catheterization. N Engl J Med 348(12):1123–1133

    Article  PubMed  Google Scholar 

  4. Climo M, Diekema D, Warren DK, Herwaldt LA, Perl TM, Peterson L et al (2003) Prevalence of the use of central venous access devices within and outside of the intensive care unit: results of a survey among hospitals in the prevention epicenter program of the centers for disease control and prevention. Infect Control Hosp Epidemiol 24(12):942–945

    Article  PubMed  Google Scholar 

  5. Institute of Medicine (US) Committee on Quality of Health Care in America. To Err is Human: Building a Safer Health System [Internet]. Kohn LT, Corrigan JM, Donaldson MS, editors. Washington (DC): National Academies Press (US); 2000 [cited 2023 Aug 1]. Available from: http://www.ncbi.nlm.nih.gov/books/NBK225182/.

  6. Randolph AG, Cook DJ, Gonzales CA, Pribble CG (1996) Ultrasound guidance for placement of central venous catheters: a meta-analysis of the literature. Crit Care Med 24(12):2053

    Article  CAS  PubMed  Google Scholar 

  7. Brass P, Hellmich M, Kolodziej L, Schick G, Smith AF (2015) Ultrasound guidance versus anatomical landmarks for internal jugular vein catheterization. Cochrane Database Syst Rev 2015(1):CD006962

    PubMed Central  Google Scholar 

  8. Mehta N, Valesky WW, Guy A, Sinert R. Systematic review: is real-time ultrasonic-guided central line placement by ED physicians more successful than the traditional landmark approach? Emerg Med J. 2013;30(5):355–9.

  9. Shekelle PG, Wachter RM, Pronovost PJ, Schoelles K, McDonald KM, Dy SM et al (2013) Making health care safer II: an updated critical analysis of the evidence for patient safety practices. Evid Rep Technol Assess (Full Rep) 211:1–945

    Google Scholar 

  10. Troianos CA, Hartman GS, Glas KE, Skubas NJ, Eberhardt RT, Walker JD et al (2011) Guidelines for performing ultrasound guided vascular cannulation: recommendations of the American society of echocardiography and the society of cardiovascular anesthesiologists. J Am Soc Echocardiogr 24(12):1291–1318

    Article  PubMed  Google Scholar 

  11. Practice Guidelines for Central Venous Access (2012) Anesthesiology 116(3):539–573

    Article  Google Scholar 

  12. Bouaziz H, Zetlaoui PJ, Pierre S, Desruennes E, Fritsch N, Jochum D et al (2015) Guidelines on the use of ultrasound guidance for vascular access. Anaesth Crit Care Pain Med 34(1):65–69

    Article  PubMed  Google Scholar 

  13. Liu C, Mao Z, Kang H, Hu X, Jiang S, Hu P et al (2018) Comparison between the long-axis/in-plane and short-axis/out-of-plane approaches for ultrasound-guided vascular catheterization: an updated meta-analysis and trial sequential analysis. TCRM 14:331–340

    Article  Google Scholar 

  14. Gao YB, Yan JH, Ma JM, Liu XN, Dong JY, Sun F et al (2016) Effects of long axis in-plane vs short axis out-of-plane techniques during ultrasound-guided vascular access. Am J Emerg Med 34(5):778–783

    Article  PubMed  Google Scholar 

  15. Lv Y, Liu H, Yu P, Wang G, Liu M, Li Y et al (2019) Evaluating the long-, short-, and oblique-axis approaches for ultrasound-guided vascular access cannulation: a systematic review and network meta-analysis. J Ultrasound Med 38(2):347–355

    Article  PubMed  Google Scholar 

  16. Convissar D, Bittner EA, Chang MG. Biplane Imaging Using Portable Ultrasound Devices for Vascular Access. Cureus. 13(1):e12561.

  17. Panidapu N, Babu S, Koshy T, Sukesan S, Dash PK, Panicker VT (2021) Internal jugular vein cannulation using a 3-dimensional ultrasound probe in patients undergoing cardiac surgery: comparison between biplane view and short-axis view. J Cardiothorac Vasc Anesth 35(1):91–97

    Article  PubMed  Google Scholar 

  18. Convissar D, Bittner EA, Chang MG (2021) Biplane imaging versus standard transverse single-plane imaging for ultrasound-guided peripheral intravenous access: a prospective controlled crossover trial. Crit Care Explor 3(10):e545

    Article  PubMed  PubMed Central  Google Scholar 

  19. Scholten HJ, Broens G, Meesters MI, van Houte J, van den Broek RJC, ter Horst L et al (2023) Biplanar versus conventional two-dimensional ultrasound guidance for radial artery catheterisation. BJA Open 1(5):100122

    Article  Google Scholar 

  20. Jones NA, Simmons CJ, Castañeda P, Carius BM, Cronin AJ, Monti J. Impact of a novel biplane user interface on ultrasound-guided vascular access performance: a prospective, randomized, crossover study. Med J (Ft Sam Houst Tex). 2023;(Per 23–4/5/6):25–30.

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

JR: collected and analyzed the information on each procedure and contributed to the writing of the document. GZ: performed the analysis of the collected information. LP: collected and analyzed the information on each procedure and contributed to the writing of the document. YB: collected and analyzed the information on each procedure and contributed to the writing of the document. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yury Forlan Bustos Martínez.

Ethics declarations

Ethics approval and consent to participate

The present is a simulation study; therefore, no patients were recruited, and it was approved by the ethics committee of the Universidad del Rosario DVO005 1932-CV 1553 on May 17, 2022. All specialists agreed to participate under informed consent.

Consent for publication

All participants signed informed consent forms for participation and publication of the results approved by the institutional ethics committee.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions